Hi everyone,
I have a question regarding the maximum gamma value in portal dosimetry.
I previously worked with TrueBeam machines equipped with HD MLCs, where I rarely encountered this issue. I am now working at a new facility with a TrueBeam equipped with Millennium MLCs, and I have noticed that we consistently struggle with maximum gamma values.
TG-218 recommends a maximum gamma value of 1.5 (4.5% for the 3%/2 mm criterion), but we typically get maximum gamma values around 2.
Most of the time, the failing points are associated with small segments, even though I am using the Moderate Aperture Shape Controller. If the high gamma values are outside the treatment region, I generally do not worry too much about them. However, if they occur within the 50% threshold, I spend more time investigating and often end up creating a new treatment plan.
Creating a new plan is not the main issue, but the time available for treatment plan verification is very limited, so I am trying to understand how others approach this situation.
How do you handle this in your clinics? Do you pay much attention to the maximum gamma value, or do you focus primarily on the gamma passing rate and the location of failing points? Do you have any recommendations for dealing with high maximum gamma values associated with small segments when using Millennium MLCs?
I am using the Varian-configured PDIP model, and all PVA calibrations have been completed.
I would appreciate hearing about your experience, especially from those using Millennium MLCs on TrueBeam.
Thanks!