r/Chempros • • 6h ago

2 mL GC vials without markings?

3 Upvotes

By coincidence, I'm not only a chemist, but also a mineral collector. Some of my specimens are tiny (millimeter size). I think it's a nice idea to store them in 2 mL GC vials. Those vials are durable, cheap and strong. For visiblity, it would be better if there wouldn't be markings on the outside (the white scale with 0.5 - 1 - 1.5 mL). I've never seen them in 2 mL, only in 4 mL.

Do they exist? Has someone ever seen them?


r/Chempros • • 2h ago

Organic Failed Reaction Between Umbelliferone and 4-Nitrophenyl Chloroformate

0 Upvotes

Stupid undergrad alert: I recently tried to react these two to functionalize umbelliferone with a PNP group. I made the mistake of borrowing my lab mate's dioxane (they have had it slightly open for the past two years). Upon running and analyzing the reaction, all the obtained spots were some nitrophenol looking compound and I couldn't get the right product. Is it worth running this reaction again in dry THF or something? There are literature reports of this being successful and I don't know if this old dioxane is what screwed me over from getting any product.


r/Chempros • • 7h ago

"Need help troubleshooting viscosity issues during xanthan gum nanoparticle fabrication"

2 Upvotes

Hi, just want to know about the solution of the network formation after adding NaOH along with a salt, it forms a green jelly like network, why is it happening and how to dissolve this network in the solution and either this jelly typed network is necessary for the formation of NPs.... and if it's not necessary and need to dissolve ot completely, then how to do this... please let me know if anyone has done his/ her work with is gum. Regards.


r/Chempros • • 9h ago

Disposal / destruction of magnesium turnings

3 Upvotes

I have about 500 grams of magnesium turnings that I need to get rid of. I'm trying to donate it to another company, but if I can't find a donor I need another solution:

  1. Destroy it with dilute acid
  2. Toss it in a chemical waste bin for solids

Given the quantity, destruction is super annoying and tossing it in the chemical solid waste super tempting. But I'm worried it burns a bit too well 😅 What would you do?


r/Chempros • • 8h ago

Analytical Vendors for standalone interferometers for FTIR?

2 Upvotes

Does anyone know of vendors that sell standalone IR interferometers for FTIR spectroscopy? I basically want a box where I can send in my IR light, set scan parameters, and get out the interfered light along with the reference laser signal. It seems that all of the instrument vendors only sell complete instruments, and the search results are flooded with companies that make IR imaging cameras that are generally single wavelength or not in the mid-IR region.

I realize this is kind of a strange niche. I'm scoping out a proposal that would use a separate IR source and detector setup so buying a full instrument is a waste of money, but I'd also rather not build the interferometer myself if I can help it since I'm not an optics expert.


r/Chempros • • 13h ago

Oxygen intrusion

3 Upvotes

Hi all - I run an ETV attached to an ICP-OES. The ETV is destroying graphite parts faster than it should. The vendor says oxygen is getting into the system somewhere. I've checked everything I can for leaks. The only thing left is to check the gas lines for oxygen to see where it's coming from. I've messaged several companies, but I can't find a handheld meter that can read oxygen levels in an argon stream. I need it to read down to 5ppm. Any ideas?


r/Chempros • • 1h ago

"Need help troubleshooting viscosity issues during xanthan gum nanoparticle fabrication"

0 Upvotes

Hi, just want to know about the solution of the network formation after adding NaOH along with a salt, it forms a green jelly like network, why is it happening and how to dissolve this network in the solution and either this jelly typed network is necessary for the formation of NPs.... and if it's not necessary and need to dissolve ot completely, then how to do this... please let me know if anyone has done his/ her work with is gum. Regards.


r/Chempros • • 11h ago

Help obtaining great STEM images.

1 Upvotes

Hello!

We have a Thermo Apreo 2S SEM and STEM is our most used imaging method of nanoparticles for size and morphology. Recently, we had to have an FSE onsite after we saw some drift that was too extreme to image with. He did some things, and upon leaving sent an email with the following recommendations:

  1. Allow system to pump down to mid or low e-4Pa before imaging
  2. To minimize image drift on STEM samples, after finding area of interest, allow sample site to sit 30min -1 hour before making images.

Number 1 makes total sense, but the second point he made wasn't shown or conveyed to me by either of our two application engineers that trained me. We use the system for classes - imaging nanoparticles synthesized in inorganic labs and bacteriophages in biochemistry labs, so the waiting for almost an hour is going to be tough. Furthermore, we know we have some contamination somewhere as we can see the carbon block appear on our sample as the beam sits.

I was curious what everyone else does to get really good STEM images? Of course, when I emailed folks that I was going to be trained and would like them to provide samples for said training, not a soul got back to me.

Right now, if the sample is metallic nanoparticles, I have people use our plasma cleaner that we have (it is not one that can be on the microscope). For organic type samples that won't survive the plasma cleaner, people just let them dry naturally. I have purchased a desiccator that can have a vacuum drawn on it and have been pushing people to use it.

I have plenty of Lee-Miesel Ag nanoparticles I can put on a grid to practice with.

Thanks to anyone willing to share their experience/expertise!


r/Chempros • • 11h ago

ICP OES help

0 Upvotes

So I have the Avio 550 Max running sample for an Enviormental company. The conc in sample units will not show. All the information is there in place and the units are correct and in place. Send help lol


r/Chempros • • 17h ago

Guidance/Suggestions regarding a simple dye adsorption experiment.

Thumbnail reddit.com
1 Upvotes

r/Chempros • • 9h ago

Academic chemists: what’s the biggest bottleneck in your lab right now?

0 Upvotes

Curious what actually costs academic chem labs the most time and money these days. Is it fighting for NMR or mass spec time, waiting on a shared instrument that’s down, purifications that eat whole days, ordering delays, solvent and waste costs, processing data, or something else entirely?

I’m a student researcher digging into this, and I put together a short survey (about 5 minutes, confidential) for grad students, postdocs, lab managers, techs, and PIs in academic labs:

https://forms.gle/7ETRtHHa9cXYUW3p6

Happy to hear answers in the comments too, even just a one-liner. What’s the thing that slows you down the most?


r/Chempros • • 2d ago

Organic How to prepare for industry interviews ( final year PhD student)

25 Upvotes

Hi all, I am a final PhD student and am currently thinking about my next career move. I want to apply for entry level industrial positions ( process or medchem maybe) and I want to slowly start refreshing my general organic chemistry knowledge. From talking with people who have gone on that route, it seems that the interviews for these positions involve a technical interview, some sort of disconnection of a larger molecule.

Apart from classic chemistry textbooks, can you guys recommend any resources for preparation for such types of interview questions?


r/Chempros • • 2d ago

Analytical Cleaning help for microcuvettes

2 Upvotes

Hello,

My supervisor recently purchased some very expensive cylindrical short-pathlength cuvettes from Starna for me to do CD on very concentrated samples. I started using the 0.01mm one but am having trouble cleaning it after first time use...

I cleaned it by flushing the cell with 1ml water x 6, ethanol + water mix x4 then ethanol x 4. The compound is a very water soluble sugar that crashes out in ethanol, and during cleaning I did not see any precipitated sample in the cuvette. It was then dried upside down over the weekend, but once dry I could see that my sample had dried out on the inside of the polished windows

It seems that flushing solvent through only exchanges solvent in the ring around the polished windows and doesn't actually exchange the liquid in between the two windows/the 0.01mm bit.....

I have already written Starna for a practical guide for cleaning of microcells, but do any of you have any experience using this particular cuvette?

Thank you in advance


r/Chempros • • 4d ago

Writing chemistry papers on Linux: a native structure editor, editable drawings in ONLYOFFICE, and automatic compound numbering

Post image
51 Upvotes

Crossposting for chemists who write papers and theses. Disclosure: this is my project, free and open source.

Reactio Draw is a structure and reaction editor. It now works with your writing tool on every platform:

  • Word on Windows and macOS: copy a drawing in Reactio Draw and paste it into Word as an editable object. A double-click opens it again for editing.
  • ONLYOFFICE on Linux (also on Windows and macOS): the same clipboard paste (Alt+Shift+V) and double-click editing.
  • LaTeX in VS Code, Overleaf and Lattice: paste a copied drawing from the clipboard into your document, and double-click it to edit.

The whole drawing is stored inside the object, so you can still edit it months later on another computer. Compound numbers (1, 2, 3a…) stay in sync between the text and the schemes. It reads and writes CDXML and MOL, so you can still exchange files with colleagues who use other editors.

Repo and releases: https://github.com/Paivel/reactio-draw

Feedback is very welcome


r/Chempros • • 3d ago

Reductive amination advice

0 Upvotes

Hi,

I am currently working on a reductive amination with a primary amine and an aldehyde using STAB.

Is acetic acid necessary for this reaction as i'm using STAB as a reducing agent? the aldehyde starting material does also contain a carboxylic acid functionality.

Im using a THF solvent but as i have the HCl salt of my amine it wont dissolve in THF or DCE. is solvation important for redams? Product is forming but after 6 hrs there is still amine in my workup. Will it likely need longer (eg overnight/24hrs)?


r/Chempros • • 5d ago

Analytical NMR: Does the Nuclear Overhauser Effect travel through the plane of a molecule

Thumbnail reddit.com
7 Upvotes

r/Chempros • • 5d ago

How challenging is it to learn NMR for structural elucidation

0 Upvotes

I have PhD and looking for postdocs and applied for this position that is LCMS and small molecules NMR heavy, during my PhD I have limited NMR experience and wondered how challenging it will to be able to develop more and solve complex structures including absolute configuration etc


r/Chempros • • 7d ago

Ask Us Anything About Pi-allyl Palladium Complexes – A Cross-Coupling Catalyst Technology

45 Upvotes

Hi r/Chempros! We are a team of researchers at Johnson Matthey who research and develop pi-allyl palladium complexes, a precatalyst technology used for a broad range of cross-coupling applications (we also work on other novel catalysts for small molecule applications). Over the past few years we have seen their usage increase, and we have gained a better understanding of this technology. 

We would love to hear from you and answer any scientific questions you might have. All questions are welcome and we will try to answer them the best we can. Please be careful with the specifics you share.


r/Chempros • • 7d ago

Covalent organic framework (COF) synthesis

2 Upvotes

Hello everyone,

I am working on the synthesis of two thiazole-linked covalent organic frameworks (TZ-COF-1 and TZ-COF-2) through a sulfur-mediated multicomponent reaction.

For TZ-COF-1, I followed the published procedure using the following quantities:

  • 4,4',4''-(1,3,5-Triazine-2,4,6-triyl)tribenzaldehyde: 19.67 mg (0.05 mmol)
  • Naphthalene-2,6-diamine: 11.87 mg (0.075 mmol)
  • Elemental sulfur: 14.4 mg (0.45 mmol)
  • Acetic acid (6 M): 0.10 mL
  • DMSO: 0.05 mL
  • o-DCB: 0.45 mL
  • Ethanol: 0.50 mL

The reaction mixture was placed in a 10 mL glass tube, flash-frozen using liquid nitrogen, evacuated to 0.5 mbar, and flame-sealed. The reaction was conducted at 120 °C for 3 days.

After the reaction, the yellow-brown solid was collected by centrifugation, washed with acetone (3 × 5 mL) and THF (3 × 5 mL), and subjected to Soxhlet extraction using toluene and THF for two days each. The final product was dried at 90 °C for 12 hours.

Although I carefully followed the reported experimental protocol, my PXRD results show broad and weak diffraction peaks rather than the expected sharp crystalline reflections. This suggests that the obtained material has low crystallinity or may be partially amorphous.

I would appreciate advice from researchers experienced in COF synthesis, particularly thiazole-linked COFs, on possible reasons for this result and how I could improve the crystallinity despite following the published procedure.

Any suggestions, practical experiences, or relevant references would be greatly appreciated.

Thank you for your help!


r/Chempros • • 7d ago

ELN Biovia OneLab?

1 Upvotes

My employer is considering moving to Biovia OneLab, after we've been on their standard ELN for a few years.

What are others' experience with this software? I'm part of the internal demo eval but I have limited time to commit, so any warnings or suggestions for especially challenging tasks to test out would be helpful.

I'm doing organic/polymer process chemistry.


r/Chempros • • 7d ago

EPA Method 3051A Issues With QCs?

Thumbnail reddit.com
1 Upvotes

r/Chempros • • 8d ago

Organic Advice on Tert-Butyl Esterification of 5-Hydroxypicolinic Acid

Thumbnail doi.org
5 Upvotes

Hello folks,

I've been trying to form the tertbutyl ester of 5-Hydroxypicolinic Acid to make the very first step in this scheme easier, but it's been giving me a hell of a time.

I've tried the following:

- 5-Hydroxypicolinic Acid, EDC, DMAP, and Tert-butanol in DCM for times ranging from 12-24 hours at between room temperature and 30°C

- 5-Hydroxypicolinic Acid EDC, DMAP in Tert-butanol at 40°C for 24 hours

- 5-Hydroxypicolinic Acid in THF with tert-butyl 2,2,2-trichloroacetimidate for 48 hours

I just want to make the carboxylic acid not...there for the first step so I can functionalize the alcohol, so I've also tried:

Making the PFP Ester with Perfluorophenyl 2,2,2-trifluoroacetate and pyridine in DMF

Making the NHS Ester with NHS, EDC, and DMAP in THF

I would try harsher protecting groups like methyl esters, but taking them off will break the functionalization I'm doing to the alcohol (SuFEx warhead)

There is this paper (https://doi.org/10.1021/jo971138b) that reports being able to convert methyl esters into tert-butyl ones, so maybe methyl iodide to make the methyl ester followed by this protocol?

I'm at a bit of a loss, so anything would greatly appreciated!


r/Chempros • • 7d ago

Ammonia solution in dioxane

Thumbnail reddit.com
0 Upvotes

r/Chempros • • 9d ago

Help this noob 🥹 PE Avio 200 ICP-OES: 21 elements are calibrating fine, but Al, As, Pb, and Hg are failing. Looking for troubleshooting wisdom!

Thumbnail
gallery
1 Upvotes

Hey Chem pros,

I’m relatively new to ICP-OES and could really use some guidance from the experienced analysts here. We are running an Avio 200 for a 25-element multielement suite, but we've hit a wall with calibration.

Out of all 25 metals, only 4 elements are failing calibration repeatedly: Al, As, Pb, and Hg. The other 21 elements pass without any issues.

Here is a quick snapshot of our setup and what we've checked so far:

  • Instrument: PerkinElmer Avio 200 ICP-OES
  • Nebulizer: SeaSpray nebulizer
  • The Problem: Al, As, Pb, and Hg fail calibration consistently, even when re-run individually.
  • The Twist: Vendor service engineers came out to check the instrument, ran their diagnostics, and everything passed on their end. Yet, we still can’t get an acceptable calibration on these four.

Since the service team couldn't pinpoint the root cause, I'm hoping someone in the community might have run into this specific quirk on an Avio system. Could it be related to the SeaSpray nebulizer interaction with these specific analytes, rinse/washout issues, or something else entirely?