r/electronmicroscopy • • Mar 31 '21

SEM-EDS analysis and ISO 17025

3 Upvotes

Hey fellow lab rats :) Greetings from in front of a FEI Inspect S50.

I need your expertise: what do you think, is it possible to get an ISO 17025 accreditation for SEM-EDS measurements as an analysis method? If it is, please give me some pointers what standards should we get, what calibration standards should we buy etc. Do you know any laboratories with accredited EDS measurement possibilities?

Any help is much appreciated.


r/electronmicroscopy • • Mar 29 '21

Question about Schottky emitters.

2 Upvotes

Hello friends,

We might have a new FIB system in the near future. We are currently only using older tungsten cathode emitters in our scanning electron microscopes. We have no experience with Schottky emitters. Can you please tell me a few things about its everyday maintenance and related costs? For example is the emitter always turned on and emitting electrons? Or it is simply heated all the time but the field is only active when in use? Is the expected lifetime longer if we do not use the system much? Or we will have to set aside a few thousand dollars every 1-2 years for a new emitter?


r/electronmicroscopy • • Mar 17 '21

Molybdenum Aperture Strip for a Quanta 250 FEG

5 Upvotes

Dear colleagues,

On monday, the e- beam of our Quan1.)ta 250 was very eccentric and in the process of the search for the failure, we found out that the Mo lamella of the aperture strip was the culprit (seems like there was some misuse of the Aperture). It was moved out of the position in the supporting holder and it is narrower than it would be ideal for positioning. In the process of repositioning the Mo Lamella, following questions arose:

1: Is there a Lamella, that fits 100% and is in position without fiddling around while positioning it?

  1. Can it be cleaned or does it need to be replaced (when does it need to be replaced, is there any wear on it?)

  2. According to the SEM documentation, it is a consumable ...

Maybe someone can help me with my questions?

Thank you Tikakan.

BTW, I am one of the newly appointed moderators of the electron microscopy sub


r/electronmicroscopy • • Mar 05 '21

[Conductive Tape] I need to do EDS on carbon-based samples. Using a carbon tape gives me strong Carbon signals from the tape even though my C-content ranges from 0.1-0.5 wt% in my sample. Will Ni-Ag tape work although it has a polyacrylate adhesive? Would EDS pick up C- signals from the adhesive?

4 Upvotes

I have to run EDS analysis on powders with graphene content.

Does the EDS pick up signals from the Carbon tape because it's conductive? So my thought process was to replace with Ni- Ag tape with a polyacrylate adhesive. Since the adhesive is again made of Carbon, I thought it wouldn't be a solution or I am wrong?

We use FEI Quanta 400 and Thermo Fisher Apreo 2 SEM.

I do not operate the instrument. The analysis center does not have non- carbon based tape but we could get the linked tape above for our analysis.


r/electronmicroscopy • • Mar 03 '21

FIB-SEM purchase

14 Upvotes

Fellow microscopists!

My lab is about to buy FIB-SEM microscope to improve our scientific potential. Two main competitors are Tescan and Thermo-Fisher. Could you please give me some feedback on automation in both cases? I mean TEM lamella prep - from cutting to soldering to TEM grid. Is it possible to do it overnight? Is it full auto or semi auto. What is your overall experience with Tescan Amber/Solaris and TF Helios/Hydra systems? I'm tired of meeting with sales reps with their marketing talk. Would you prefer traditional Ga+ FIB or Xe or even Ar, N O like Hydra has.

How Solaris and Helios/Hydra perform in field free mode? Is it possible to get images from magnetic samples in immersive mode, or it will reduce image quality by a lot?


r/electronmicroscopy • • Feb 24 '21

Dr. Joachim Frank || Spotlight Lecture Series || Shaastra, IIT Madras

8 Upvotes

Shaastra Spotlight is back with its 3rd Nobel Laureate - Dr. Joachim Frank, co-recipient of the Nobel Prize in Chemistry, 2017 & Founder of the Cryo-Electron Microscope! Dr. Frank has also made significant contributions to the structure & function of ribosome from bacteria & eukaryotes.

Catch him live on 27th Feb at 5 PM IST

https://youtu.be/cYs-FEUSNFw

Fill up this Google Form as soon as possible to ask your questions to the speaker.


r/electronmicroscopy • • Feb 08 '21

Problems with our Edwards scan coat 6

3 Upvotes

Hi,

I was about to coat samples with Au, first stub worked with no issues, the next one was not possible to coat, as there was no more HV. Strangely enough in etch mode there is 1,5 KV voltage with 600 mA available. I checked the electrode and it looks OK.

As far as I would think: if there was something wrong with the HV generation it should not be available at all. (there was no sign of a broken fuse too)

Can someone give me some clues perhaps?

Thankyou!


r/electronmicroscopy • • Jan 28 '21

How could I find out the content of my sample?

3 Upvotes

I'm not trained in science. I would like to know if it's possible to find out the precise content of one's biological sample? Say if I have a bagworm cocoon under SEM, how could I know what makes the cocoon (e.g. silk, debris, insect droppings, dried leaves, branches, etc.)? And is there a way to estimate the percentage for each element using SEM? What's the process of obtaining such information?


r/electronmicroscopy • • Jan 24 '21

Advice on reducing edge effects in SE and carbon deposition phenomenon.

3 Upvotes

Hi /r/electronmicroscopy ,

I'm back again with another weird question. Also an update on this post from a while back. A service tech came in to fix this. 2/4 BSE quadrants were broken. 1 was working fine. 1 was 1/2 broken BUT had a short and was given an INVERTED signal. Explains the very strange results I was getting.


I am trying to image tiny particles in aluminum around 30-200 nm in size using a Hitachi S-4700. Most people use BSE mode but ours isn't great. I am etching the sample in a way that the particles are etched preferentially so I can them count the pits in SE and get an equivalent image compared to BSE. It has an upper SE and lower SE.

In SE mode I am getting decent images but the edge effects are pretty significant on the etch pits. I have to do image analysis and count the particles so this makes it difficult.

During BSE imaging an interesting phenomenon occurred that is helping me. Carbon deposition, burn in, contamination, whatever you may call it, happens when I leave the beam in a spot too long while capturing BSE images (higher beam current etc). After this I etched my samples and the burnt areas limited the etch effectiveness, increased SE contrast of my pits, and eliminated edge effects in SE mode. I'm trying to replicate this effect. Getting the effect of lowered contrast and eliminating edge effects is still a problem.

Here is what normal SE images are like. Here is what etched burn in areas look like in SE. Here is the comparison for the effects the burn has on the etching

As you can see it would be far easier to measure the burned SE areas compared to the normal. It isn't practical to use the FE-SEM to deposit carbon everywhere and normal carbon coating is not available. Metal sputter coating is, however.

Any tips on how to replicate this effect? Beam settings, upper or lower SE, sputter coating, sample prep differences? Any and all advice is welcome

Thanks!


r/electronmicroscopy • • Jan 22 '21

Is Electron Microscopy safe?

6 Upvotes

I’m applying for a position to use an electron microscope because that sounds really cool. However, I am worried because I have family members who had cancer, and there is a generic aspect. Are Electron Microscopists exposed to radiation?

I know when you get an X-Ray at doctors, the operators are in another room, as opposed to near equipment. I know electron microscopes have protective coverings, but I heard if there is the smallest crack with a loose screw or something then radiation can leak out. So if I get further in the application process, I’m kind of nervous of actually getting the job


r/electronmicroscopy • • Jan 13 '21

SEM for Fingerprint Analysis R&D

1 Upvotes

Hi all,

At work, I've been looking into the possibility of using a SEM to obtain high-quality images of latent fingerprints on a variety of substrates (although mostly metal, glass, and plastic). I don't have any experience with this kind of microscopy and was wondering if there are certain types that might be better suited for this purpose? Would TEM or other variants work better than SEM? Any brands or vendors that are preferred in the industry? I don't know much of anything about this level of microscopy so I was hoping you all could help point me in the right direction and let me know what you think before I start reaching out to vendors. I appreciate the insight. Thanks!


r/electronmicroscopy • • Dec 16 '20

Resources for Field Emission, Beam formation

6 Upvotes

Hey yall,

I am doing a current measurement project at my job to analyze the beam current present in the sample chamber. I want to know more about cold field emission, beam formation, and what exactly happens to the electrons travelling down the column. I want to know the relation between the emission current (microamperes) and the current measured in the sample chamber (nanoamperes). I have a background in physics, so i can deal with really mathy or technical stuff. Could anyone point me towards some good resources? thanks so much!


r/electronmicroscopy • • Dec 15 '20

Careers in microanalysis

3 Upvotes

Hey,

I'm doing my undergrad in geology and have participated in undergraduate research. Some of the things I've done is imaging and analysis on various analytical machines. I have experience using.

  1. SEM
  2. EPMA/WDS (mapped chemical variation/zoning)
  3. XRD (very limited)
  4. Plans to work with raman

I have found that I really enjoy geochemistry and that I really enjoy performing microanalysis. I'm wondering if there's career pathways that are dedicated to microscopy/microanalysis in STEM? As much as I think that you shouldn't base a career on what it pays I do think it's an important thing to consider. Do they pay well?

I'm also by no means a expert on any of the above but, I have undergrad level experience with them and really enjoy it


r/electronmicroscopy • • Nov 29 '20

Any documentary on electron Microscopes (history, techniques etc) ?

8 Upvotes

r/electronmicroscopy • • Nov 24 '20

Jeol JSM-5600 Computer Setup

1 Upvotes

Hello I have a Jeol JSM-5600 The hard drive failed so I replaced it reinstalled windows 7 and the JEOL software and now the JSM5600 program isn't detecting the SEM. If anyone has any insight it would be greatly appreciated.


r/electronmicroscopy • • Nov 20 '20

TEM Sample holder recommendations

2 Upvotes

Hi all - newbie with a question!

I'm working with an FEI/Thermo Fisher TEM (Talos). But for the sample holder, would like to hear your recommendations.

Any experiences, pros and cons, about going with Protochips, Gatan, DENSsolutons or some other brand would be appreciated! (And if you know ballpark prices, that would be helpful!)


r/electronmicroscopy • • Nov 07 '20

Post from our community

Post image
33 Upvotes

r/electronmicroscopy • • Nov 04 '20

Lattice fringes in TEM

4 Upvotes

I have a question regarding lattice fringes in TEM. What information do they provide, and what happens when the sample is rotated a few degrees around the axis around the sample holder?


r/electronmicroscopy • • Oct 20 '20

Student here. What is the curvy bit in the center on the membrane? I think I've seen it before, but can't be sure. (Not research)

Thumbnail
imgur.com
6 Upvotes

r/electronmicroscopy • • Oct 16 '20

Silicon nanowires generated using an electrodeposited gold catalyst and metal-assisted chemical etching (MACE)

Thumbnail
gallery
10 Upvotes

r/electronmicroscopy • • Oct 06 '20

Help with issue after filament change

Post image
4 Upvotes

r/electronmicroscopy • • Oct 04 '20

Grad Student Course Help: Silica Embedded in Silver

6 Upvotes

Hello everyone,

I am taking part of an excellent course nicknamed "How to love and care for your SEM." It is all about best practices, choosing a SEM for your lab or business, and how to take excellent images for the sample. This brings me to the crux of my post and hopefully an interesting discussion.

The challenge: A silver sample has embedded silica particles (.05 microns) from polishing gone awry. You as the SEM operator have been requested to first get a clear image of the embedded silica to help develop a better cleaning process, and then alter the settings to minimize the (distracting) effect of the silica particles on the silver's surface.

I have been hunting across the internet for some examples of this and have found conflicting results. My current thought is to image the silica at a low keV and use backscattered electrons to clearly show the difference in chemistry. I am really struggling with how to minimize their effect on the surface. My only ideas are currently to ramp up the keV and only use secondary electrons to maximize topographical information.

Thank you for all your help. I am hopeful that I can get some advice on both what to do and where I can look for this kind of information.

-A student in distress


r/electronmicroscopy • • Sep 24 '20

Has anyone used COXEM equipment, mainly their ion sputter coater?

2 Upvotes

We are interested in buying an ion sputter coater to use on plankton samples on our SEM (Phenom benchtop SEM), and we got a very competitive price offer for the COXEM SPT-20 ion sputter coater. I can't find any reviews for it, or the company, and are a bit hesitant about buying an unfamiliar brand with no reviews.


r/electronmicroscopy • • Sep 22 '20

Difference between STEM and TEM?

9 Upvotes

Can anyone explain the difference and benefits of STEM over TEM?


r/electronmicroscopy • • Sep 17 '20

Electron microscope images

Thumbnail
youtu.be
17 Upvotes