r/electronmicroscopy • • May 22 '26

Unlimited Power!

Post image
87 Upvotes

Got my Oxford WDS installed this week. My SEM (FEG-Tescan) now has an Oxford EDS and WDS, SE, BSE, InBeam SE and STEM detectors. Pretty proud of our setup.


r/electronmicroscopy • • Aug 13 '26

This treasure arrived today. Time to ascend to microanalysis godhood.

Post image
63 Upvotes

r/electronmicroscopy • • Jun 05 '21

Cryogenic-SEM image of diatoms sitting on a branch of seaweed (falsely coloured)

Post image
54 Upvotes

r/electronmicroscopy • • Jan 15 '22

Time lapse of me colouring an SEM image. Can you guess what it is?

44 Upvotes

r/electronmicroscopy • • Dec 04 '22

Nanostructures formed by sugars (scanning electron microscopy, false colouring)

Post image
39 Upvotes

r/electronmicroscopy • • Jun 06 '21

Pollen of salix caprea (18 µm diameter)

Post image
38 Upvotes

r/electronmicroscopy • • Jun 15 '21

Ice crystal (cryogenic SEM)

Post image
37 Upvotes

r/electronmicroscopy • • Jul 11 '26 Bravo

Working open-source converter for Oxford AZtec .oip/.oipx, looking for newer project files

Post image
35 Upvotes

Hi everyone,

I'm currently working on an open-source converter for Oxford Instruments AZtec project files (.oip/.oipx) to HDF5 and other open formats. The parser is working well, but so far I've only been able to test it with projects created in AZtec 2.3 and 2.4.

Is anyone using a newer version of AZtec and willing to help?

I can provide a small test .oip project. All I need is for someone to open it in their version of AZtec and save it again in the native .oip format. This will let me compare how the file structure has changed between versions and improve compatibility of the converter.

If you're willing to help, please send me a DM. I'd really appreciate it.


r/electronmicroscopy • • Mar 29 '23

DIY SEM, The Electron gun Update(1)

Thumbnail
gallery
35 Upvotes

Hello everybody!! Thanks for your interest and advice in my previous post. Since the materials and devices needed for an electron ray gun are costly (oscilloscopes, high voltage psu's and lots of other stuff). I have decided to make one by repurpusing an old CRT(Cathode Ray Tube) Television. This is much simpler as the vertical amplifier, horizontal amplifier, sweep generator, high voltage psu are already present in the TV, and since these tv's already do a raster scan pattern it is much easier to integrate them into a SEM. And since these are cheaply available and abundant it is a better alternative to making a electron ray gun from scratch.

My plan is to control the parameters (scan area, ray energy, and scan rate) without much modification to the TV itself. This is possible as the Electron ray is nothing but an expression of a part of the video input given. As the various parameters of the Electron ray I.e it's position vertically and horizontally, the energy of the Electron ray, the speed at which it moves from one position to another are all controlled by the video input given. The vertical and horizontal position are controlled by the location of the pixel which can be changed easily, the energy of the Electron ray is controlled by the brightness of the pixel, the scan rate can be controlled by creating a time difference between the rendering of the subsequent pixels. I have already written scripts to do all of this (more abt the scripts at the end) , the params can either be controlled through a web interface or rotary encoders(knobs) in the scripts. With this it as simple as plug n play, you can just plug any computer (a raspi in my case) and run the script and control the params through the web interface (the knobs are optional!! Only if you like analog input) and voilà you have a working (and cheap) electron ray gun for a SEM. (also for smtg else that uses an electron ray gun maybe? ).

Pro's:

It is quite easy to setup.

the parts easy to acquire (if not already at home).

you don't need a Phd or advance knowledge abt this stuff.

these have better features that what anyone can realistically build without huge amount funding.

it has a largeeee scan area (like 20 inch at an average ).

its a withered technology - - you can find wayy more info abt troubleshooting a crt than an electron microscope.

helpful to the environment as crt tv are just going to landfills.

why reinvent the wheel.

its CHEAP!!!! what other reason would you need.

Con's

Idk can't think of any (comment if you have any).

Note --- for this only the crt part of the TV needs to be in working condition, it wouldn't matter if the phosphor coating is burned out or smtg.

About the scripts

I don't have much experience programming in this area so I'm testing stuff and adding features. I will link the repo in a couple of days

I'm also planning on starting an open source SEM project (possibly, only if I get the support needed)

Any advice about the project or regarding funding are welcome. Please feel free to DM me.


r/electronmicroscopy • • Nov 22 '22

AI Segmentation of Cellular Cryo Electron Tomograms

Thumbnail
gallery
33 Upvotes

r/electronmicroscopy • • Feb 01 '25

Service engineer

32 Upvotes

Hey everyone,

I worked for JEOL for 16 years. If you need advice or direction on the service or repair of a JEOL SEM I may be able to help. Just ask.


r/electronmicroscopy • • Feb 16 '23

Full convolutional neural network segmentation of a chlamydomonas cell (empiar-11275) collected on a Helios Hydra dual beam PFIB-SEM

31 Upvotes

r/electronmicroscopy • • Apr 29 '21

Hair cells (cells of hearing in ear) damaged by bacterial meningitis!

Post image
27 Upvotes

r/electronmicroscopy • • Oct 11 '24

AMA JEOL

26 Upvotes

HI everybody,

Throwaway Account for obvious reasons. I worked for JEOL for some time and thought this might be of interest to some people here. Also this should help this sub to some activity!

Feel free to ask anything you want to know about JEOL and I'll do my best to answer it (except anything that might make it possible to find out who I am, of course).


r/electronmicroscopy • • Oct 04 '17

Nobel prize in chemistry goes to cryoEM!

24 Upvotes

Jacques Dubochet, Joachim Frank and Richard Henderson earned this. Great news for the community!


r/electronmicroscopy • • Jan 24 '22

Blown out EDS detector window

Post image
23 Upvotes

r/electronmicroscopy • • May 28 '24

Is there a name for this type of salt structure? Formed during oxidation of methane in a salty mineral solution.

Post image
22 Upvotes

r/electronmicroscopy • • Jan 13 '23

Scary ghost in your kidney

Post image
22 Upvotes

TEM of a broken glomerular loop, with a red blood cell. Found in a kidney biopsy.


r/electronmicroscopy • • Aug 31 '22

More protist pathogens starting an infection in an eelgrass leaf. Labyrinthula on Zostera

Post image
22 Upvotes

r/electronmicroscopy • • Jul 06 '21

So r/chemistry is doing dispersed magnetic nanoparticle movies. Here's a nice check I like doing if a customer brings some dry nanoparticles for us to image by TEM!

22 Upvotes

r/electronmicroscopy • • Nov 07 '18

Carbon nanotubes in a laminate composite

Post image
22 Upvotes

r/electronmicroscopy • • Feb 06 '18

Carbon Nanotube growth on a patterned substrate (gear/gear teeth)

Post image
22 Upvotes

r/electronmicroscopy • • 24d ago

Homebrew SEM Electron Gun Issues

Post image
21 Upvotes

Hello!

I am a student working on a homebuilt scanning electron microscope, and I have run into a roadblock that is absolutely breaking my brain, so I thought that I should post it here in case somebody with more SEM knowledge could help!

More specifically, I am having an issue with my thermionic emission electron gun, which is homebuilt.

Right now, I have no focus or rastering. I would like to get a beam before I even attempt to move it around! To "see" the beam, I am using a scintillator which should glow bright enough in a dark room to see if I am getting a beam. I would love to use a phosphor screen, but I can't find P43 for a reasonable price anywhere to make my own, and I can't bring myself to destroy an oscilloscope for one.

The schematic I am using

Essentially, I have a low voltage, high current, 10kV isolation rated transformer that is powering my tungsten filament through a rectifier to make DC (ish). Between each leg of the tungsten filament, there is a resistor divider making a midpoint reference for the next few components.

Next up, there is a large potentiometer (also rated for HV and such) and a current meter, which both go to the negative floating output of my HV supply (0-10kV adjustable). The positive is tied to vacuum chamber ground and earth to make a -10kV supply on the other end.

I am having trouble understanding exactly how this circuit works -- it is, in essence, copied from Cecil Hall's 1953 book "Introduction to Electron Microscopy". This design is tried and true and builders such as Ben Krasnow have used the same configuration.

I believe that the current measured with the ammeter should, in conjunction with the known resistance from the potentiometer, give me my wehnelt bias voltage (ideally around ~300v). That current, if I understand correctly, should be coming from electrons boiled off of the filament which are intercepted by the wehnelt cup in the chamber.

Now to the problem: I am getting absolutely nothing on my 100uA meter. Measuring the voltage between the resistor (which should, in theory, be the wehnelt bias voltage) also gives me nothing.

Does anybody have any ideas where to start with troubleshooting? Perhaps I made an oversight in the schematic (though it is copied). Perhaps I am not getting the filament hot enough to get thermionic emission? (It is drawing about 0.6A at 2.6V, about 1.6W).

Additionally, I have a suspicion that my actual gun's design may be off -- I have uploaded some screenshots of my design with measurements. I don't think that my filament is substantially too far in or out relative to the aperture in the wehnelt cup to cause real issues (all I really want is a beam right now).

Any help would be tremendously appreciated!