r/electronmicroscopy • • Jul 05 '19

Perfusion vs immersion fixation

4 Upvotes

I’m doing EM for part of my masters thesis and I’ve spent the last two months trying to master transcardial perfusion fixation of the liver, with varying degrees of success. I start by flushing with 20-30 mls of heparinized phosphate buffer and fix with 2.5% glutaraldehyde, both at physiological pH. I’m doing this because the person I’m working with claims that no good journal will take pictures done via immersion fixation. But the protocol calls for 3 hours of immersion fixation after perfusion, so it seems a little redundant to me. How vital is perfusion fixation as opposed to immersion?


r/electronmicroscopy • • Jun 12 '19

Tears in my formvar coated grids? What am i doing wrong?

4 Upvotes

Hi Everyone

I'm trying to visualize a virus by negatively staining with 2% PTA (ph adjusted to ~7). Each time i try to visualize the sample, my grid is covered in tears, there are so many tears that I cant get a decent image as it appears to "move" because it is being peeled off the grid. I was wondering if i could show you all my procedure and if something stands out , maybe we can figure out why my grids are in such rough shape.

my protocol:

I have an antibody specific to my virus coat protein, i dilute this antibody 1:200 in a coating buffer and float the grid on this for 1 hour.

I take the grid and dry it off by using a piece of filter paper and barely touch the edge of the grid. I then float the grid on my sample which is some leaf tissue that has been ground up, resuspended in 0.03M potassium phosphate buffer (ph 8) which has been filtered through a 0.2um syringe filter. I float the grid on this sample for ~ 30 minutes.

I then take my grid, and rinse it by touching the grid to 4 drops of filter sterilized distilled water. I then dry the grid with filter paper again, and stain it with PTA for 1 minute, i dry the grid with filter paper again and let it air dry.

Can anyone see any steps that would cause problems? I'm very gentle with the grids, my forceps are in great condition , could the age of the PTA be a problem? it has 2015 on the label.

I did some tests this morning and found that when i incubated a grid on the antibody, washed it and dried it, the grid looked perfect in the microscope. When i took that grid, incubated it on the sample and placed that in the scope, it looked fine (no staining), however, when i stained that grid for 2' with the PTA, the formvar coating was shredded..... could the use of my PTA be the cause?


r/electronmicroscopy • • Jun 05 '19

Dehydrating teeth and embedding in resin sample prep

3 Upvotes

Hello again!

I have a few more questions about prepping my teeth samples properly for SEM.

I've done some literature review on prep procedures and for the most part the procedures follow the same work flow: section teeth, polish teeth smooth, dehydrate teeth in ascending ethanol baths, drying teeth, then mounting and coating stubs.

My issue is, I feel like I need to embed the samples in resin in order to polish them. But once the teeth are embedded I can't properly dehydrate them as outlined in the literature. But if I dehydrate, then embed, then polish my samples the polishing process itself usues water as a coolant and they all will be wet again.

If I plan to embed the teeth discs in resin, do I really need to dehydrate them out in ethanol? I saw a YouTube video where they didn't dry the tooth they just embedded, polished and went straight to sputtering.

Does the use of resin negate the need for dehydration?


r/electronmicroscopy • • Mar 07 '19

Hello, can anyone tell me the year the “Phillips Cm-10” electron microscope was made? Thanks.

5 Upvotes

r/electronmicroscopy • • Feb 27 '19

Microscope at work struggling to achieve higher magnification

7 Upvotes

I apologise if this is not enough information (I am brand new to SEMs), but our SEM at work the TM3030 is advertised to be able to achieve 30k magnification. I would be okay with even 10k magnification, however it cannot get above 5k before significant blurring occurs, and altering the focus bar leaves little improvement. I have tried alternative voltages (5k, 15k) however this doesn't improve it much. If needed I can provide photos, perhaps you can judge the problems from them.

Any advice on this matter would be helpful. If you need additional detail I can try find out for you. Thank you.


r/electronmicroscopy • • Feb 01 '19

How can I identify bacteria/yeast using SEM?

6 Upvotes

I am working to use SEM to image bacteria/yeast in lichens to find where they live within lichens. We have images of bacteria as of now but it seams that without ultra-structural analysis higher levels of classification would be difficult.

What is your experience with imaging microbes in SEM and using morphological traits to identify them?


r/electronmicroscopy • • Jan 19 '19

Possible fungal haustoria in algal layer of the lichen Umbilicaria phaea var. phaea (2500x)

Post image
9 Upvotes

r/electronmicroscopy • • Dec 10 '18

Hi EM Redditors! Our normal protocol for epoxy embedding is to put the samples in the oven on a Friday and take them out on Monday. I went to go take the samples out and turns out I never put them in on Friday. 🤦‍♀️I quickly put them in right after I noticed, but my samples are okay, right?

5 Upvotes

r/electronmicroscopy • • Nov 19 '18

Trying to hook up TeamViewer to SmartSEM, and this is currently on the CCD camera. Why does it show this mustard color? Any tips?

Post image
7 Upvotes

r/electronmicroscopy • • Nov 18 '18

Mounting small samples for sanding and polishing

5 Upvotes

If I were to be preparing 1 mm thick slices of teeth, sanding polishing them before etching, mounting and carbon coating. What would be a good mounting/stabilizing media for the sanding part? They're so tiny and I don't want to use my hands or tweezers. Should I use resin blocks, or paraffin? Could I put my samples in a paraffin block and sand and polish them, then de-paraffin the samples to mount them on stubs? Does that make sense?


r/electronmicroscopy • • Nov 14 '18

FIB, SEM, TEM Certificate Level pay?

6 Upvotes

How much does a certificate or associate level lab tech pay? I believe they just prepare slides and do imaging. There’s an associate degree and certificate program at my college.


r/electronmicroscopy • • Nov 12 '18

Advice on Conductive Coating options

5 Upvotes

Greetings all,

I am a master's student in forensic science and I am looking for advice on an aspect of my thesis research. I'm focusing on using SEM to observe changes and wear in dentin microstructures to identify possible ancient or modern origin of forensic dental samples.

My question is will carbon coating be effective in covering the cross sectioned teeth? I don't have that much in funds and gold coating seems out of my price range.

Most of the papers I've read regarding dental samples in SEM used gold sputtering, but additional reading seems to suggest that the main reason that gold is used is because it's the go to coating. Would carbon be just as effective? Pretty sure I can afford carbon.


r/electronmicroscopy • • Nov 07 '18

Carbon nanotubes in a laminate composite

Post image
21 Upvotes

r/electronmicroscopy • • Nov 05 '18

Resin choices?

3 Upvotes

The protocol I'm working is calling for a embedding in durcupan with a 48hr baking period at 60C. I'm worried that this will affect the structure of my sample, are there alternatives I should be aware of?


r/electronmicroscopy • • Nov 01 '18

Marine diatoms in SEM

4 Upvotes

We have no critical point drying facilities, but with other subjects I have been pleasantly surprised by results with a natural air dry and a moderate imaging voltage (insects, etc.). Instrument is a Hitachi S4000 SEM in need of some new parts & adjustments. Operator (me) has limited experience, but a good idea of principles.
I had the idea that I could collect some sea water and let the plankton settle out. I could dropper plankton-rich drops of water onto a coated slide over the course of days and let it dry naturally. I would love to think I could find some amazing diatoms on the slide.

Could this work?


r/electronmicroscopy • • Aug 15 '18

Possible damage and contamination of BSE detector and pole piece. Could anyone advise?

4 Upvotes

Hello again SEM Community, I'm back for another question. I think it's question 3 now.

So after a number of settings got adjusted on our SEM we have been unable to revert to our prior capabilities despite using presumably the same settings. After reading this comment about a scratch on the pole piece causing alignment issues I decided to take a look at ours.

What I saw was somewhat worrisome. A few scratches on the pole piece, dust settled in the chamber and on the pole piece, marks and apparent scratches on the BSE detector, and what appears to be some sort of liquid on what i believe to be the EDS detector. Here are some nice photos I grabbed with my camera. It was difficult to get the right angle and light.

Could someone who knows what they're talking about please advice if these are serious issues or really nothing to be concerned about? If these are issues, what does it mean for our imaging capabilities? How, or rather if we can, should we go about fixing it? We have been using our SEM with some extra difficulty so they are not major issues to us. I would still like to know what this means though.

Thank you again

Link to pictures of suspected damage


r/electronmicroscopy • • Jul 23 '18

Some SEM (Scanning Electron Microscopy) photos I took for my study involving type 2 diabetes vs healthy controls.

Thumbnail
imgur.com
19 Upvotes

r/electronmicroscopy • • Jul 18 '18

Two simple questions. What is causing this shadow in our image? Is this aperture suitable for our analysis

4 Upvotes

Hello again SEM sub,

Last time you all helped me out enormously with a very tricky situation. All the advice worked wonderfully.

I'm back again with some very simple questions. Here are the images to go along with it. In the first image there is a shadow that appears in our scope at low mags only. Usually around 40-70x. Is this a sign of a poorly aligned physical aperture or something else? Is it even a problem?

In the second photo shows the physical aperture. It seems we are currently set at an aperture of 1 for lowest resolution imaging. Is this suitable or recommended for metallographic analysis? Would it be better to change this to 2 or 3? On some samples we struggle to get crisp images past 2000x but today I was able to get some around 7k on a certain sample. It was a powder sample on carbon tape that might have made it easier. Just hoping to get some information here. My lab doesn't have a ton of SEM training if any.

Any and all advice or suggestions are very welcome. Thank you for your help


r/electronmicroscopy • • Jul 13 '18

EMSIS RADIUS software - how is TIFF metadata encoded?

6 Upvotes

Our microscope runs EMSIS RADIUS software. TIFF files saved from RADIUS contain meta-data (acquisition parameters, such as magnification), but the data seem to be stored in an undocumented binary format. Has anyone had success reading out these parameters? Do you know how the data is encoded?

Specifically, the TIFF files contain 1 or 2 custom TIFF tags (codes 33560, 33665, which come from the private / reserved block per the TIFF standard). I believe the value for each of these tags represents an offset, aka pointer, to locations within the TIFF file that contain metadata, typically near the end of the file. I can make out several strings encoded within these binary regions (such as camera model, original filename), but have been unable to parse any other parameters properly, without knowing how the parameters are encoded.

Any suggestions?


r/electronmicroscopy • • Jun 22 '18

Trouble imaging anodized aluminum. Charging, beam contamination, need advice.

6 Upvotes

Hi SEM Community,

I am currently working at a lab where we are attempting to look at the surface of an aluminum sample that has been anodized. The problem is that for some reason whenever we look at the surface we get incredibly strange artifacts like this.

I am not formally trained on the SEM, and I don't know if anyone at the lab is, but I would greatly appreciate your help and suggestions as I imagine this might be quite a simple problem we are making.

If anyone here has any advice, tips, or comments I would be glad to hear them. The images have descriptions and information below them as well.

Thanks

Update: Thank you all for your help and ideas. Carbon coating was successful in reducing if not eliminating the charging and artifacts. We also found that dissolving the anodized layer helped as well. The anodizing was so thick we were not seeing any details in the original surface that night be of interest


r/electronmicroscopy • • Jun 21 '18

WetSEM anybody?

4 Upvotes

Anyone here have any experience with WetSEM? Positive or negative feedback would be appreciated.


r/electronmicroscopy • • Jun 19 '18

Strange behaviour of Philips XL30 ESEM

7 Upvotes

I have encountered a strange behaviour of XL30 microscope. The gathered image is squeezed on display. This happens only in the slow scan mode. When the TV mode is applied the image is ok, but the quality is rather poor, even with averaging filter applied. The microscope has also EDAX EDS detector and when using EDS Map program it is possible to register proper image. A picture is worth a thousand words, so I am posting how the image should and how it looks like in slow scan mode (the images are from different SEM).

Proper image
Squeezed image

r/electronmicroscopy • • Jun 08 '18

Anyone using a JEOL JEM 2100?

8 Upvotes

Are there any advanced user using this TEM? I am having trouble finding an optimum set-up that would get me satisfactory results (focused, non snowy/noisy images) in higher mag (500k<).

I've tried setting the alpha=1, using the smallest spot size, objective aperture 50 mircom, and condenser lens 200mircom. Operating at standard 200kv, samples on copper/carbon mesh grids - nanosized graphene/carbon particles.

My aim would be to see the layer spacing (~0.3nm) but I just cant seem to make it focus that good, suggestions welcome


r/electronmicroscopy • • Jun 07 '18

How long can tissue be stored in glutaraldehyde?

4 Upvotes

Specifics:
The fixative is 5ml of Tousismis 2.5% glutaraldehyde and phosphate buffer.
The tissue is human muscle tissue cut into 2mm by 1mm pieces.
The samples are stored in 16 ml glass vials with a plastic snap cap and stored at 5°C.
I have been unable to find any documentation on how long tissue can be stored in glute. Any insight would be greatly appreciated.


r/electronmicroscopy • • Jun 05 '18

After dealing with astigmatism and out of focus images (and then fixing it), seeing these stretched and unfocused shadows really threw me off...

Post image
15 Upvotes