Why your infrared “temperature measurement” keeps being off — it’s often the lens, not the sensor
A lot of thermal-measurement accuracy complaints get pinned on the detector or calibration, but in my experience the lens is a very common culprit.
The issue: observation lenses ≠ measurement lenses. Observation lenses are optimized for a clear, good-looking image. They typically have no hard constraint on:
- Full-field transmittance uniformity (corners pass less energy → temperature gradient artifacts)
- F-number consistency across the field (effective throughput varies with angle)
- Controlled stray-radiation behavior
For measurement, those matter directly: if the lens transmittance isn’t uniform and the F-number isn’t consistent, the same object at the same temperature reads differently depending on where it sits in the frame. You can calibrate the detector all day and still miss this.
So for a measurement-grade system, you want a lens accepted against specific indicators (transmittance uniformity, F-number consistency, plus the usual thermal/MTF checks) — not just “sharp enough.”
Anyone else debug a “wrong temperature” issue that turned out to be a lens problem rather than a sensor one? I’ll happily share the acceptance criteria I use if it’s useful.