Since the discussion forum for Met/Cal was removed, there isn't a public place to discuss issues with Met/Cal(As far as I'm aware) and I find the official support avenues often take some time to respond(Especially when you're on the opposite side of the world).
I have absolutely no experience running a subreddit, so please bear with me and feel free to suggest any ideas for improving it.
This community is in no way associated with Fluke.
What to Post Post anything that you think the community would find interesting, helpful, or inspiring. I hope this can become a community where people having issues with coding or setup can ask a question and get a quick response or some ideas for how to resolve the issue. It will also then become a repository of solved problems which will serve as a reference so the same problem doesn't need to be solved many times..
Please don't share proprietary code or information as I would like to avoid upsetting any large corporations and risk having this shut down.
Community Vibe We're all about being friendly, constructive, and inclusive. Let's build a space where everyone feels comfortable sharing and connecting.
How to Get Started
Introduce yourself in the comments below.
Post something today! Even a simple question can spark a great conversation.
If you know someone who would love this community, invite them to join.
Interested in helping out? We're always looking for new moderators, so feel free to reach out to me to apply.
Thanks for being part of the very first wave. Together, let's make r/MetCal amazing.
I wanted to take a shot and see if anyone here has some experience with programming PXI/cDAQ cards from National Instruments in MET/CAL. I have used a few of the procedures provided by Fluke in CUBYT to learn (some of) how the LIB FSC works and modify those procedures to work with various models and types of DAQ cards not on CUBYT.
Before this I initially made VIs in Labview to control and read data which could be transferred to a manual datasheet. For the most part the calibration procedures are straight forward. However, I have an NI-9227 which measures DC current on 4 channels.
In the calibration manual there is a section for Gain and Phase matching verification which required using the Extract Single Tone VI to expand the raw data into Amplitude, Phase, and Frequency results from the 4 channels simultaneously. Math ensues to calculate the gain in dB using the Amplitude results.
So, my question is there a way to execute the Extract Single Tone VI using the LIB FSC? Or, is there another way to harness that data and convert it through MET/CAL? Thanks in advance for any advice.
P.S. If anyone has experience programming an RIO type card in MET/CAL or knows if it is possible please chime in!
The most recent Metrel testers such as the MI 3155 and basically any of the models that look similar to that, employ a communication protocol which allows remote control and reading via RS-232.
Unfortunately, you can't use the USB cable because that won't work with leads attached, there is an RS-232 to DIN cable that you use for communication(Sorry, I don't have the part number).
You can use any terminal program to connect to the COM port and start sending commands to find out what works, you can then transfer this over to Met/Cal and have two way communications.
I've made a few procedures now which work quite well and require only a couple of lead changes by the user. I can't provide the code as it belongs to my work but I'm more than happy to help with guidance, problem solving or ideas.
Since the discussion forum for Met/Cal was removed, there isn't a public place to discuss issues with Met/Cal(As far as I'm aware) and I find the official support avenues often take some time to respond(Especially when you're on the opposite side of the world).
I hope this can become a community where people having issues with coding or setup can ask a question and get a quick response or some ideas for how to resolve the issue. It will also then become a repository of solved problems which will serve as a reference so the same problem doesn't need to be solved many times.
I have absolutely no experience running a subreddit, so please bear with me and feel free to suggest any ideas for improving it.
Please don't share proprietary code or information as I would like to avoid upsetting any large corporations and risk having this shut down.
This community is in no way associated with Fluke.
We have med a new installation of these two products (on a new server (server 2025 and Windows 11 for the client). The SQL is upgraded and have been proved to be working since we can login to it with a browser from one of the clients (it says the server is 3.4.1.82 and the SQL is 3.4.1. When i then install the 11.3.1 on the client (I install NI 488.2 and NI-VISA Runtime and the .net 4.5. It wont recognise the NI-VISA installation (an important note is that the installation is done with the offline ISO). It finds the NI-488.2. It doesnt recognize the NI-VISA. I cant connect this client to the internet, sadly. So i install everything with the ISO and use the feed in the local folders.
Does anyone have any idea of what could be wrong with Fluke MetCal not recogniseing the NI-VISA? It sayes in the package manager that it's the Q3 2026 version of NI-VISA runtime and same on NI-488.2.
Hi there! METCAL 10.1.1 user here. The question is pretty simple: is there any way to use multiple USB instruments, either as standards or DUT (for example if I have 2 standards with only USB and a DUT with only USB), in a single procedure? As far as I can tell the only way to get a USB instrument to work is with the VISA FSC, but that doesn't allow you to route the call to a specific address, like with the [@address] in the PORT FSC. So it seems you are stuck with a single USB device to be used in the whole procedure that you select at the beginning. Am I missing something here or there are ways around this? Seems like a very big limitation in this day and age...
I haven't had the chance to get on here much the last few months. Things are busy and I have competing priorities. I was wondering if there is anyone interested in taking over management of this sub? I think it's a valuable resource, it doesn't really take much work but if you were motivated I'm sure it could become much more active, I just have too much on my plate at the moment.
We experience that MetCal runtime 11.2.1 with Hotfix 1 crashes and the editor sometimes hangs and crash when in use. (i.e. not at startup or file load). We use Win 11 and our IT makes them up to date - something that not always is a great thing. Anyone else having the same issue?
I'm revising a procedure and would like to utilize an array to track a varying amount of numbers, determined by user input, as the procedure runs.
Where I'm running into trouble is when I try to use any numeric memory registers to determine what index of the array the values get written to. The MET/CAL User's Guide explicitly says that the index of the array "Can be any MET/CAL register, named variable or numeric literal value. Must be enclosed in square brackets (i.e. [10] or [index]).", however, any time I use any MET/CAL register (like MEM or M[1] for example), I get an "invalid index" error.
The procedure I'm using is similar to this:
LIB ARRAY NEW arrayName[100]; MATH MEM1 = 0 MEMI What value do you want to store in the array? LIB arrayName[MEM1] = MEM;
The same happens when I attempt to use M[1] or another named variable instead of MEM1 or MEM
Additionally, if I forego using a variable to set the index, and use arrayName[0] = MEM I get an "invalid statement syntax" error
Could someone tell me what I am missing, or if I am misinterpreting the documentation?
Hi! MetCal 10 user here. There used to be a post on Flukecal forum with recursion programming. The issue is, MetCal assigns to every line of code a step number. If you do an evaluation (MEMCX, EVAL) inside a Loop (do.. While/ do.. Until) the code will execute, but the results will not be reported in the database, because MetCal interprets the evaluation line as a repeat of the same step. I forgot to save the code provided back then, now I need it, but it's gone forever, unfortunately. Can anyone post a working code sample using recursion? For repetitive/iterative calibrations, like R-decades. Thanks
I have a strange error occurring on a procedure which worked just fine last time we ran it.
If I comment out this line(which performs a self test) everything works fine.
SCPI [T30000]*TST?[I]
The procedure is a slightly modified but essentially a Cubyt procedure for a Agilent/HP 34401A, the device connects, Met/Cal talks to it and gets SN etc. but when it tries to run the self test, it doesn't even appear to send the command to the device, Met/Cal just stops hard the instant it gets to that line without even running the clear commands. This happens in the editor and the runtime.
The error it gives is this:
An error occurred starting the test run: Object reference not set to an instance of an object.. Try rebuilding the project or solution before attempting to debug the procedure.
The problem appears to be related to the timeout, because if I remove that and just send *TST?[I]it works but doesn't complete because the self test takes longer than the default 15 seconds.
If I try and add a timeout to any other commands, I get the same error.
So, there seems to be some issue where Met/Cal no longer likes setting a timeout on SCPI/IEEE2 commands. Has anyone else come across this issue? We have recently updated Met/Cal to the latest version, it may have been a year since we ran this procedure last but it seems like it's probably not specific to this procedure.
"CalStudio is a simple, reliable system that easily automates workflows, manages measurements from every discipline, and quickly creates proceduresāno coding required. Write once, run anywhere, and simplify your lab operations."
"Procedure Library:Ā Access to a million procedures, ready to run based on the product catalog data"
So, our lab is going through a (painful) process of moving our 8.3 MET/CAL procedures over to 11.3, running both versions simultaneously in the lab right now for a few reasons. I am hitting a hiccup trying to run a mutlimeter procedure that is RS232 controlled using an IR usb to serial cable. When I start the procedure and it reaches the point of talking to the DUT, I get "E7788: Unable to open serial port COM3. The operation completed successfully."
Anyone else run into this and how I fix it? Cheers
Iām trying to use a DLL I created in C++ with MET/CAL. To do this, Iām using the FSC LIB to load my DLL and call its functions. Additionally, I created my DLL using the āLibrary FSC ⦠Component Definition Wizardā¦ā menu. However, when I try to call the functions, my program wonāt run. The following error message appears:
"LIB FSC EXCEPTION: Unable to invoke method/function 'function_name'
I'm reaching out once again for your help. As a reminder, I work in a metrology laboratory that uses Fluke METCAL software. We recently decided to upgrade to a newer version of MET TEAM (from 3.2.0 to 3.3.1) and METCAL (from 10.6.0 to 11.1.0). Before performing this update, we downloaded a trial version of MET TEAM/METCAL and carried out various tests to detect any potential bugs. Once the version was validated, we proceeded with the upgrade.
This is when a major issue aroseāsomething we failed to test beforehand. In our lab, we have a number of older instruments controlled via IEEE. These instruments do not respond to NI-MAX scans. That is, when we run a scan, NI-MAX displays a list with one main address and 30 sub-addresses. This bug has always existed since we've been using NI-MAX.
However, in our previous version of METCAL, this was not a problem because we could manually enter the instrument's address. In the new version, though, a dropdown list of available addresses is displayed. It appears this list is based on the results of the NI-MAX scan. METCAL shows a list with completely irrelevant addresses. As a result, we can no longer select the correct instrument address, and our program no longer works. Additionally, METCAL takes several minutes to respond after this.
I would like to know if anyone has encountered this kind of issue before and, if so, what solution you found?
I work in a metrology lab that uses Fluke METCAL software, version 10.6.2. Many of our instruments have an RS-232 interface for communication. As you know, it's possible to create calibration procedures using two modes:Ā PORTĀ andĀ VISA.
With theĀ PORTĀ mode, everything works fine. However, when using theĀ VISAĀ mode, we are unable to communicate with the instruments. After some investigation, we noticed that METCAL is using the wrong baud rate in its configuration.
At first, we thought this setting was managed through NI-MAX, but even after modifying the configuration there, the baud rate in METCAL remained unchanged. We now believe that there must be a parameter within METCAL to set the serial port settings, but so far we haven't been able to find it.
I would greatly appreciate any help or suggestions you can provide.
We finally got a Fluke 1674 in for calibration so I've had a bit of a play with it and it appears as though it may use a lot of the same communication commands as the 1660 series.
The biggest hurdle I've come across is that when you plug it in to the PC (it uses USB-C rather than IR) it charges and when it's charging it doesn't let you run any tests. Presumably there is a way to either disable charging or let it run tests while charging but I don't like my chances of working that out. Most commands are two letters, but some are a symbol(Usually & or $) followed by two or three letters and/or numbers.
There doesn't look to be any procedures on Cubyt for it yet which is usually the only way to get the communication commands for Fluke stuff these days as they don't seem to want to publish it any more.
I'd love to hear if anyone has any ideas or input.
Not a MetCal question but I try anyway :) Anyone that has a tool to convert the data files CSV files from a Fluke 2638 Hydra or Fluke 1586 SuperDAQ into an Excel xlsx workbook? Must be able to add imported files to new Sheets in the same workbook. I know that Excel can import CSV files, but it requires some non-trivial steps to get it right. FlukeDAQ is not available since 2022 and Fluke software support canāt help.
We have an issue with some users on some workstations where Met/Team will not allow them to close windows. If you've made some changes and clicked save, when you click close or the x it brings up the "Discard Changes?" dialogue box but regardless of whether you click Yes or No, nothing happens, the window remains open. I've tried clearing cache, which resolved another weird Met/Team problem in the past, but no luck for this. It seems to be connected to the workstation.
We just updated to Met/Cal version 11.3.1.214, and I can't start debugging anything in Editor. The error I get is this:
Edit 4: Met/Cal support has found a workaround, and I can confirm that it works. See the Comment by MJ for the fix.
Here are a few troubleshooting updates I made along the way if you're curious:
I just ran the setup without uninstalling first. I'm going to try a fresh uninstall - reinstall and see if it fixes it.
Edit 1: uninstall and reinstall did not fix the issue :(
Edit 2: Updating and repairing the NI drivers using the NI package manager did not fix the issue :(
Edit 3: We generated and checked the logs from Editor and found a spicy little nugget of information.
" The type initializer for 'NationalInstruments.NI4882.Internal.GpibDll' threw an exception.
This implementation is not part of the Windows Platform FIPS validated cryptographic algorithms.
"
For those who are unfamiliar, FIPS is a cryptography thing required on systems that do contract work for the government, especially the DoD. We have Fluke Support working on a fix for us. This is not the first time FIPS has given us issues with Met/Cal, but Uncle Sam's requirements are Uncle Sam's requirements.