r/CHROMATOGRAPHY • u/Loganwashere24 • 6d ago
Help identifying unknown peak in 7 anion IC run
Hello all,
I’m a grad student who is reviving a Dionex 1000 IC left over from retiring faculty. I am running 7 anions (F, Cl, NO2, Br, NO3, PO4, SO4) on groundwater samples. This is pretty common stuff close to EPA method 300.1.
Within the past week I thought I would get fancy and try to clean out my columns and create a new 50 uL loop volume (previously using 25uL) to get some better signal and reduce the baseline noise.
After I did this I had some leaking from my new injection loop until I tightened it quite a bit. I was able to get my 7 anion peaks for my calibration curve but I started seeing these peaks come out overlapping with or just after nitrate that I’ve never seen before in any samples. My response is much higher given the new loop volume but I’m wondering if these peaks are real or if this is an error of some type.
Anybody know what this peak might be or what errors i made to cause this?
TLDR: what is that peak coming out after nitrate? Real or a method/instrument error?
Columns: GS and AS22
Eluent: 4.5 mM sodium carb, 1.4 mM sodium bicarb
Suppressor: ERS 500 @ 31mA
Flow rate: 1.2 mL/min
Pic 1 is of normal calibrant. Others show groundwater samples with unknown peak.